Getting Started - BeaconMedaes MedPlus TotalAlert Operation And Maintenance

Alarm network
Table of Contents

Advertisement

Getting Started

Chapter 2: Troubleshooting Procedures

Getting Started

WARNING:
Only facility-authorized personnel should troubleshoot the MedPlus
2
TotalAlert® Alarm Network. Troubleshooting by unauthorized personnel
could result in personal injury or equipment damage.
Begin each procedure in this chapter with step 1. Follow the sequence outlined
(each step assumes the previous step has been completed). In each step, the nor-
mal operation of the product can be confirmed by answering Yes or No to the
statement. Your response will lead to another step in the procedure, a repair
analysis procedure (RAP), or a component replacement. If more than one compo-
nent is listed, replace them in the given order.
To begin gathering information about the problem, start with Initial Actions.
To verify the repair after completing each corrective action (replacing or adjust-
ing a part, seating a connector, etc.), perform the Function Checks.
To verify the repair, perform the Final Actions after the Function Checks.
If the troubleshooting procedures do not isolate the problem, call
BeaconMedæs Technical Support at 1-888-4MEDGAS (1-888-463-3427) for
assistance.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Page 2-3