2.10 Area Alarm Module ROM Error
1. An Area Alarm Module displays error code A 05.
Yes
2. Area Alarm Module is connected properly.
Yes
3. Area Alarm Module is working correctly.
Yes
4. If the problem still exists, call BeaconMedæs Technical Support.
2.11 Battery Low on Master Alarm
1. Battery Low displays on the Master Alarm.
Yes
2. Check the Master Alarm's battery voltage. It is approximately 2.9 V DC.
Yes
3. If the problem still exists, call BeaconMedæs Technical Support.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
No
Go to "Final Actions" on page 2-7.
No
Connect the Area Alarm Module properly (refer to procedure
4.9 on page 4-26). If this solves the problem, go to "Final
Actions" on page 2-7. Otherwise, go to step 3.
No
Replace the Area Alarm Module (refer to procedure 4.9 on page
4-26). If this solves the problem, go to "Final Actions" on page
2-7. Otherwise, go to step 4.
No
Go to "Final Actions" on page 2-7.
No
Replace the Master Alarm's battery (refer to procedure 4.3 on
page 4-12). If this solves the problem, go to "Final Actions" on
page 2-7. Otherwise, go to step 3.
Area Alarm Module ROM Error
Chapter 2: Troubleshooting Procedures
2
Page 2-19