Configuring the Area Alarm by Using the Front Panel
Chapter 1: Introduction
Returning to Normal Operating Mode
1. To return to normal operation and save any changes that you have made, do
one of the following:
•
•
If you do not press any keys for 10 minutes, the alarm returns to normal opera-
tion and saves any changes that you have made.
2. While
Alarm Module is monitoring by pressing the Increment key or the
Decrement key. For a list of available gases, refer to table 1-31 on page 1-
118.
NOTE:
The type of gas or vacuum being monitored cannot be changed.
Display Shows
oy9n
M7Ed
UAcu
WJA9
nito
nit9
cArb
hELi
Oycd
Page 1-118
Press the configuration button (CE) or (CF) (see figure 1-58 on page
1–116).
Turn the power supply switch (DR) to the OFF position, and then back
to the ON position (see figure 1-57 on page 1–115).
GASt
is being displayed, you can view the type of gas that the Area
Table 1-31. Area Alarm Module Gas Types
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Gas type
Oxygen
Medical air
Vacuum
WAGD (waste anesthetic gas disposal)
Nitrous oxide
Nitrogen
Carbon dioxide
Helium
Oxygen and carbon dioxide mixture