Area Display/Transducer Fault - BeaconMedaes MedPlus TotalAlert Operation And Maintenance

Alarm network
Table of Contents

Advertisement

2.13 Area Display/Transducer Fault

1. Area Display/Transducer Fault displays on the Master Alarm.
Yes
2. The Area Alarm Module is connected to the Master Alarm.
Yes
3. The transducer is connected to the Area Alarm Breakout P.C. Board.
Yes
4. The Area Alarm Module is operating properly.
Yes
5. The transducer's LED is flashing at a slow steady rate.
Yes
6. If the problem still exists, call BeaconMedæs Technical Support.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
No
Go to "Final Actions" on page 2-7.
No
Connect the Area Alarm to the Master Alarm (refer to procedure
4.9 on page 4-26). If this solves the problem, go to "Final
Actions" on page 2-7. Otherwise, go to step 3.
No
Connect the transducer to the Area Alarm Breakout P.C. Board
(refer to procedure 4.7 on page 4-22). If this solves the problem,
go to "Final Actions" on page 2-7. Otherwise, go to step 4.
No
Replace the Area Alarm Module (refer to procedure 4.9 on page
4-26). If this solves the problem, go to "Final Actions" on page
2-7. Otherwise, go to step 5.
No
Replace the transducer (refer to procedure 4.20 on page 4-50). If
this solves the problem, go to "Final Actions" on page 2-7.
Otherwise, go to step 6.

Area Display/Transducer Fault

Chapter 2: Troubleshooting Procedures
2
Page 2-21