While Browsing the Master Alarm, Area Alarm Modules Are Not Accessible

Chapter 2: Troubleshooting Procedures
2.36 While Browsing the Master Alarm, Area Alarm Modules Are
Not Accessible
1. While you are browsing the Master Alarm, you cannot use the jump link to
go to Area Alarm Modules.
Yes
2. Type in the word Area followed by its serial number in your Internet brows-
er's address window, such as Area26453. The Area Alarm cannot be browsed.
Yes
3. The Ethernet is connected properly.
Yes
4. Check the LEDs on the Area Alarm Network Communications Board. The
LEDs that indicate whether the alarms are communicating are flashing at a
steady rate.
Yes
5. If the problem still exists, call BeaconMedæs Technical Support.
Page 2-42
No
Go to "Final Actions" on page 2-7.
No
If this solved the problem, go to "Final Actions" on page 2-7.
Otherwise, go to step 3.
No
Ensure that the Ethernet is connected properly on the Area
Alarm Network Communications Board (refer to procedure 4.25
on page 4-68). If this solves the problem, go to "Final Actions"
on page 2-7. Otherwise, go to step 4.
No
Call your facility's Information Systems personnel to fix the
problem. If they fix the problem, go to "Final Actions" on page
2-7. Otherwise, go to step 5.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)