Power Supply Fuse

Chapter 4: Removal, Replacement, and Adjustment Procedures

4.15 Power Supply Fuse

Tools required:

Removal

1. Depending on which alarm you are replacing the fuse in, take one of the fol-
lowing steps:
Page 4-40
Screwdriver
5/32" hex wrench
or
Phillips head screwdriver
Unless the replacement is being performed on a Local Alarm, use a
5/32" hex wrench to unlock the front panel (A) (see figure 4-16 on page
4–40).
When changing the power supply fuse in the Local Alarm, use a phillips
head screwdriver to remove the front panel.
Figure 4-16. Power Supply Fuse
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)

Hide quick links: