Transducer Pressure Below Sensor Range - BeaconMedaes MedPlus TotalAlert Operation And Maintenance

Alarm network
Table of Contents

Advertisement

2.1

Transducer Pressure Below Sensor Range

1. Error code A 01 displays on an Area Alarm Module.
Yes
2. The shutoff valve is open.
Yes
3. Transducer is connected to the pipeline.
Yes
4. Transducer has a slow steady flash on its heartbeat LED.
Yes
5. If the problem still exists, call BeaconMedæs Technical Support.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
No
Go to "Final Actions" on page 2-7.
No
Open the shutoff valve. If this solves the problem, go to "Final
Actions" on page 2-7. Otherwise, go to step 3.
No
Connect the transducer to the pipeline (refer to procedure 4.20
on page 4-50). If this solves the problem, go to "Final Actions"
on page 2-7. Otherwise, go to step 4.
No
Replace the transducer (refer to procedure 4.23 on page 4-62). If
this solves the problem, go to "Final Actions" on page 2-7.
Otherwise, go to step 5.

Transducer Pressure Below Sensor Range

Chapter 2: Troubleshooting Procedures
2
Page 2-13