Blank Area Alarm Module; Removal - BeaconMedaes MedPlus TotalAlert Operation And Maintenance

Alarm network
Table of Contents

Advertisement

Blank Area Alarm Module

Chapter 4: Removal, Replacement, and Adjustment Procedures

4.10 Blank Area Alarm Module

Tools required:

Removal

1. Using a 5/32" hex wrench, unlock the front panel (A) (see figure 4-11 on
page 4–30).
Turn off the power supply in the alarm, and leave it off during the installa-
tion or replacement of any component or sub-assembly. Failure to do so
could result in personal injury or equipment damage.
2. Turn the power switch (D) to the OFF position.
3. Using a phillips head screwdriver, remove the two screws (B) from the Area
Alarm blank module (C).
Page 4-30
5/32" hex wrench
Phillips head screwdriver
Figure 4-11. Area Alarm Blank Module
Shock Hazard:
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)