Incorrect Transducer
Chapter 2: Troubleshooting Procedures
2.5
Incorrect Transducer
1. An Area Alarm Module displays error code A 07.
Yes
2. Transducer is programmed for the correct type of gas.
Yes
3. Transducer has a slow steady flash on its heartbeat LED.
Yes
4. If the problem still exists, call BeaconMedæs Technical Support.
2.6
Transducer Program Errors
1. An Area Alarm Module displays error code A 10.
Yes
2. Transducer has a slow steady flash on its heartbeat LED.
Yes
3. If the problem still exists, call BeaconMedæs Technical Support.
Page 2-16
No
Go to "Final Actions" on page 2-7.
No
Connect the proper gas-specific transducer (refer to procedure
4.23 on page 4-62). If you need more information about how to
connect the transducer, refer to the MedPlus TotalAlert® Alarm
Network—Conversion Kits Installation Instructions . If this
solves the problem, go to "Final Actions" on page 2-7.
Otherwise, go to step 3.
No
Replace the transducer (refer to procedure 4.23 on page 4-62). If
this solves the problem, go to "Final Actions" on page 2-7.
Otherwise, go to step 4.
No
Go to "Final Actions" on page 2-7.
No
Replace the transducer (refer to procedure 4.23 on page 4-62). If
this solves the problem, go to "Final Actions" on page 2-7.
Otherwise, go to step 3.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)