Configuring the Area Alarm by Using the Front Panel
Tools required:
Entering the Area Alarm's Configuration Mode
To enter the configuration mode, do the following:
1. Using a 5/32" hex wrench, unlock the front panel (DO) (see figure 1-57 on
page 1–115).
2. Open the front panel (DO).
3. Locate the Network Communications Board (DP) or the Connector P.C.
Board (not shown) that is located on the back of the front panel (DO).
NOTE:
The Connector P.C. Board does not have LEDs on it.
4. Press the configuration button (CE) or (CF), located on the P.C. board until
9ASt
displays on the Area Alarm Module (see figure 1-58 on page 1–116).
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Configuring the Area Alarm by Using the Front Panel
5/32" hex wrench
Figure 1-57. Area Alarm
Chapter 1: Introduction
Page 1-115