Remote Sensor Back Box Mounting; Pipeline Connection Sensor Module - BeaconMedaes MEGA 2 Installation, Operation And Maintenance Instructions

Medical electronic gas alarm
Table of Contents

Advertisement

Rough-In Installation
Alarm Panel Back Box Mounting
Loosen screws to
adjust back box depth
(same each side)
Figure 7: Mounting Bracket Adjustment

Remote Sensor Back Box Mounting

Alarm panels ordered with remote
sensors are provided with separate metal
boxes for each sensor. Mount each
sensor box as follows:
1. Remove sensor rough-in from
mounting box (Figure 8).
2. Using four mounting holes provided,
attach box to a wall or other structural
support. (fasteners by others).
3. Reinstall sensor rough-in.
Four holes in back of box are
provided for surface mounting
Rear conduit knockouts
(either side) are recommend-
ed for sensor wiring
Figure 8: Remote Sensor Back Box
Pipeline Connection
Sensor Module
1. Braze copper extension tubes from
sensor rough-in to appropriate
pressure/vacuum piping system drops
as shown in Figure 9. Braze
connections per procedures required
by NFPA 99 or CAN/CSA-Z305.1.
Use appropriate measures to prevent
overheating and damage to internal
components of sensor rough-in
assemblies.
2. Perform standing pressure tests and
cross connection tests as required by
NFPA and CSA.
Part No. 6-847684-00 Rev. D01