Local Sensor - BeaconMedaes TotalAlert 2 Installation, Operation And Maintenance Instructions

Table of Contents

Advertisement

Finish

Local Sensor

1. Remove sensor module from shipping
carton.
2. Insert sensor into valve opening in
pressure/vacuum service rough-in
(Figure 33). An indexing pin provides
for proper orientation and prevents
gas service cross-connection. Push
sensor all the way in and secure with
(2) #6-32 x 1-1/4" screws. Repeat this
process for all sensors within alarm
panel.
NOTE:
Remove blue protective netting from sensor
connector.
3. The digital display module(s) located
on front panel assembly are factory
wired with a harness and polarized
two-pin connector for each sensor
module. Each digital display module
is labeled with the gas type. Match
gas type of each sensor to gas type of
each digital display module and join
polarized connectors of sensor and
digital display module harnesses.
NOTE:
Pressure/vacuum service rough-ins have been
installed in a specific order to optimize sensor
module wire routing.
Indexing Pin
Indexing Hole
Screws
Figure 33: Local Sensor Mounting
41
6-847718-00 Rev. B00