Operating Instructions

6.7.4. Analog I/O Configuration

Table 6-7 lists the analog I/O functions that are available in the M100E.
Sub Menu
AOUTS CALIBRATED:
CONC_OUT_1
CONC_OUT_2
CONC_OUT_3
TEST OUTPUT
AIN CALIBRATED
To configure the analyzer's four analog outputs, set the electronic signal type of each
channel and calibrate the outputs. This consists of:
Selecting an output type (voltage or current, if an optional current output driver
has been installed) and the signal level that matches the input requirements of the
recording device attached to the channel, see Sections 6.7.4.1.
Calibrating the output channel. This can be done automatically or manually for
each channel, see Sections 6.7.4.2 and 6.4.3.
Adding a bipolar recorder offset to the signal, if required (Section 6.7.4.4.)
In its standard configuration, the analyzer's outputs can be set for the following DC
voltages. Each range is usable from -5% to + 5% of the nominal range.
The default offset for all ranges is 0 VDC.
76
Table 6-7:
DIAG - Analog I/O Functions
Shows the status of the analog output calibration (YES/NO) and initiates a
calibration of all analog output channels.
Sets the basic electronic configuration of the A1 analog output (SO
three options:
Range: Selects the signal type (voltage or current loop) and full scale level of the
output.
REC_OFS: Allows to set a voltage offset (not available when RANGE is set to
CURRent loop.
Auto_CAL: Performs the same calibration as AOUT CALIBRATED, but on this one
channel only.
NOTE: Any change to RANGE or REC_OFS requires recalibration of this output.
Same as for CONC_OUT_1 but for analog channel 2 (SO
Spare
Same as for CONC_OUT_1 but for analog channel 4 (TEST)
Shows the calibration status (YES/NO) and initiates a calibration of the analog to
digital converter circuit on the motherboard.
Table 6-8:
Analog Output Voltage Ranges
Range
Minimum Output
0-0.1 V
0-1 V
-0.05 V
0-5 V
-0.25 V
0-10 V
Function
Maximum Output
-5 mV
-0.5 V
Model 100E Instruction Manual
) . There are
2
)
2
+105 mV
+1.05 V
+5.25 V
+10.5 V
045150102 Rev XB1