Setting the Frequency Range
To set the center frequency to 134 MHz, press:
Center
134
M/µ
To set the span to 30 MHz, press:
Span
30
M/µ
NOTE
You could also press the
range limits as start frequency and stop frequency values.
Setting the Source Power
To change the power level to −5 dBm, press:
−5
Power
x1
You could also press
NOTE
one of the power ranges to keep the power setting within the defined range.
Setting the Measurement
To change the number of measurement data points to 101, press:
Sweep Setup
NUMBER OF POINTS
To select the transmission measurement, press:
Trans: FWD S21 (B/R)
Meas
To view the data trace, press:
AUTOSCALE
Scale Ref

Step 3. Perform and apply the appropriate error-correction.

Refer to the
Chapter 5 , "Optimizing Measurement Results,"
correcting measurement errors.
To save the instrument state and error-correction in the analyzer internal memory,
press:
Save/Recall
SELECT DISK
and
Start
Stop
POWER RANGE MAN
or on ET models:
TRANSMISSN
INTERNAL MEMORY
Making Measurements
Making a Basic Measurement
keys and enter the frequency
POWER RANGES
for procedures on
RETURN
SAVE STATE
and select
1-5