Setting Parameters For Lc System Operation - Waters 717plus Operator's Manual

Hide thumbs Also See for 717plus:
Table of Contents

Advertisement

Table 3-2 Configuration Parameters (Continued)
Parameter
Temp Setpoint
Temp Control?
Draw Rate
Syringe Volume?
Loop Volume?
3
Installed IEEE
Address
Desired IEEE
Address

3.6.2 Setting Parameters for LC System Operation

To communicate with other LC system components:
• Set Report Delay for data integrators
• Set terminal strip signal parameters for Hold, Injection Start, and Injection Stop
• Set IEEE-488 address
Setting Report Delay
The Report Delay provides time for the data integrator to process data from a run before a
new injection begins.
If a data integrator is attached to the LC system, set the delay as follows:
1. From the 717plus Main Menu, press
63
Startup and Configuration
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Configuration
Page
2
Sets temperature for sample compartment
heater/cooler. See
Temperature
Control.
2
Enables temperature control for units with
heater/cooler. See
Temperature
Control.
3
Sets syringe draw rate. See
Injecting Viscous
3
Set equal to installed syringe size. See
Section 4.8.2, Microbore/Semiprep
3
Set equal to installed sample loop size. See
Section 4.8.2, Microbore/Semiprep
4
Displays current address. See
Setting Parameters for LC System
4
Sets new IEEE address. See
Setting Parameters for LC System
Config
Page.
Description
Section 4.8.1,
Section 4.8.1,
Section 4.8.3,
Samples.
Injections.
Injections.
Section 3.6.2,
Operation.
Section 3.6.2,
Operation.