Installing sample injection options

Installing the small-scale sample injection valve option (PrepLC
system)
If you expect to routinely inject sample volumes of less than 2-mL volume,
Waters recommends you use the small-scale manual injector valve option kit.
This kit includes these components:
Rheodyne 7725i variable-volume injector valve assembly
100-µL loop assembly with fittings
Inject start cable assembly
Small/large scale select valve with label
Fittings kit
To install the small-scale sample injection valve:
1.
Remove the two blank-off plugs from the injector panel. Pressing them
outward, from behind.
2.
Place the self-adhesive "small / large scale" label over the opening below
the existing "pump / purge" valve.
3.
Fit the 2-way select valve in the "small / large scale" opening. The valve
must rotate through both positions.
4.
Remove the injector handle from the main valve assembly by loosening
its two set screws.
2-40