Preventing contamination

For information on preventing contamination, refer to Controlling
Contamination in LC/MS Systems (part number 715001307). You can find
this document on http://www.waters.com; click Services and Support >
Support.

The selector valve

The selector valve is located on the right-hand side of the instrument, behind
the visor. The letters etched on the front of the valve represent the
components to which the associated port is connected. Etched letters with a
light background indicate inputs, while letters with a dark, etched
background indicate outputs. The following table outlines the component to
which each port connects, and whether the port is a fluid input or output.
Selector valve connections:
Port
P
A
B
R
W
LC
S
D-2
Plumbing the IntelliStart Fluidics System
Component
Infusion pump
Sample reservoir A
Sample reservoir B
Rinse/wash bottle
Waste bottle
LC column
Xevo TQD ion source
Input/Output
Both
Input
Input
Input
Output
Input
Output