Fitting the new IonSABRE II probe heater

Required materials
Chemical-resistant, powder-free gloves
IonSABRE II probe heater
To fit the new IonSABRE II probe heater:
1.
Use the probe adjuster knob to adjust the capillary so that it protrudes
approximately 0.5 mm from the end of the probe.
2.
Carefully slide the probe heater over the capillary sleeve on the probe
assembly.
3.
Fit the probe heater to the probe assembly, ensuring that the heater is
fully seated on the probe assembly.
4.
Fit the probe to the instrument (see
5.
In the Instrument Console, click API
4-100 Maintenance Procedures
Take great care not to damage the probe heater's
Caution:
electrical connections, capillary sleeve, or capillary when fitting
the heater over the capillary sleeve.
Probe heater connections
To avoid damaging the probe heater's electrical
Caution:
connections, do not twist the heater when fitting it to the probe
assembly.
0.5 mm
Capillary
page
3-7).
to start the desolvation gas.
TP02670
Capillary sleeve