Fitting The New Source Enclosure And Probe Adjuster Assembly Seals - Waters Xevo TQD IVD Overview And Maintenance Manual

Table of Contents

Advertisement

3.
Use the O-ring removal kit to carefully remove the following seals from the source
enclosure:
Source enclosure seal
Nebulizer gas seal
Desolvation gas seal
Warning:
environmental regulations. The seals can be contaminated with biohazardous or toxic
materials.
4.
Dispose of all the seals in accordance with local environmental regulations.
4.25.2
Fitting the new source enclosure and probe adjuster assembly
seals
4.25.2.1
Required materials
Chemical-resistant, powder-free gloves
Wash bottle containing HPLC-grade (or better) 1:1 methanol/water
New seals
Warning:
spreading contamination to uncontaminated surfaces, wear clean, chemical-resistant,
powder-free gloves when working with the source components.
Nebulizer gas seal
TP03164
To avoid spreading contamination, dispose of the seals in accordance with local
To avoid personal contamination with biohazards or toxic materials, and to avoid
September 7, 2016, 715003560IVD Rev. C
Desolvation gas seal
Source enclosure seal
Page 107