7.29.2 Fitting the new IonSABRE II probe heater

Required materials
• Chemical-resistant, powder-free gloves
• IonSABRE II probe heater
To fit the new IonSABRE II probe heater:
Notice:
the probe heater's electrical connections, capillary sleeve, or capillary when fitting the
heater over the capillary sleeve.
1. Use the probe adjuster knob to adjust the capillary so that it protrudes slightly from the end
of the capillary sleeve.
Figure 7–114: Adjusting the capillary
Probe heater connections
Capillary
Capillary sleeve
2. Carefully slide the probe heater over the capillary sleeve on the probe assembly.
3. Fit the probe heater to the probe assembly, ensuring that the heater is fully seated on the
probe assembly.
4. Fit the probe to the instrument (see
5. Start the desolvation gas.
The probe heater is easily damaged. Take care not to bend, crush, or distort
Notice:
To avoid damaging the probe heater's electrical connections, do not
twist the heater when removing it from or refitting it to the probe adapter.