10. Dispose of the O-rings in accordance with local environmental
regulations.
Fitting the new APPI lamp drive assembly O-rings
Required materials
•
Chemical-resistant, powder-free gloves
•
2.5-mm Allen wrench
•
1:1 methanol/water
•
Lint-free cloth
•
Mounting shaft insertion tool
To fit the new APPI lamp drive assembly O-rings
1.
Ensure that all the grooves for the O-rings are free from dirt and hairs.
If contamination is present, use 1:1 methanol/water, applied to the
Tip:
lint-free cloth, to carefully clean the grooves.
2.
Fit the new lamp mounting flange O-ring to the APPI source enclosure
side plate.
4-12
Maintenance Procedures
The O-rings may be contaminated with
Warning:
biohazardous and/or toxic materials. Ensure that they are
correctly disposed of according to local environmental
regulations.
To avoid self contamination with toxic and
Warning:
biohazardous materials, always wear chemical-resistant,
powder-free gloves at all times while handling the
components. The probe and source may be contaminated
with biologically hazardous materials.