Gripping the probe heater as shown, carefully pull it off the probe
assembly.
3.
Dispose of the probe heater in accordance with local environmental
regulations.

Fitting the new IonSABRE II probe heater

Required materials
Chemical-resistant, powder-free gloves
IonSABRE II probe heater
Caution:
connections, capillary sleeve, or capillary, take great care when
fitting the heater over the capillary sleeve.
Probe heater
To avoid spreading contamination, dispose of the
Warning:
probe heater according to local environmental regulations. It
can be contaminated with biohazardous and/or toxic
materials.
To avoid damaging the probe heater's electrical
Replacing the IonSABRE II probe heater
February 28, 2014, 715002212 Rev. C