Installing The New Capillary - Waters Xevo TQ-S Operator's, Overview And Maintenance Manual

Mass spectrometry system
Table of Contents

Advertisement

4 Maintenance Procedures

Installing the new capillary

Required materials
Chemical-resistant, powder-free gloves
Combined 2.5-mm Allen wrench and cone extraction tool
10-mm wrench
Needle-nose pliers
LC pump
HPLC-grade (or better) 1:1 acetonitrile/water
Capillary
Ferrule
Seal
PTFE liner tubing
Conductive sleeve
Red PEEK tubing
Sharp knife or PEEK tubing cutter
Metal gasket for the probe tip
Safety goggles
To install the new capillary
1.
Use the sharp knife or PEEK tubing cutter to cut an approximately
60 cm (24 inches) piece of red, PEEK tubing.
Requirement:
horizontal axis).
180
February 28, 2014, 715002212 Rev. C
To avoid personal contamination with biohazards
Warning:
or compounds that are toxic or corrosive, and to avoid
spreading contamination to uncontaminated surfaces, wear
clean, chemical-resistant, powder-free gloves when working
with the probe and source components.
To avoid puncture wounds, handle the probe with
Warning:
care; the ESI probe tip is sharp.
:
Cut the tubing squarely (that is, perpendicular to its

Hide quick links: