Removing Board Dx32 - Sirona ORTHOPHOS XG 3Dready Service Manual

Table of Contents

Advertisement

Tab 6
Removing the remaining covers
59 38 399 D3352
D3352.076.01.26.02 09.2016
6.2 Replacing the height adjustment motor (M1_4)/spindle
Remove the top cover and pull connector X607 off board DX1.
Now remove the following cover parts:
– Arm cover, top
– Slide cover, top rear
– Slide cover, bottom rear
– Slide cover, front
Note: FH laser cable (PAN)!
i
NOTE
To prevent scratches, also remove the covers of the slide table and the drawer.
6.2.3

Removing board DX32

Removing board DX32
Tip: In order to facilitate removal of board DX32, the bottom edge of the
slide (H) should be roughly flush with the upper bracket in the stand.
1. Unscrew the bracket (C) located in front of board DX32.
2. Remove the cable ties (A) from cable L2.
6 – 11