6 Cleaning and maintenance
6.8 Removing and cleaning the drain tank
86
3. Clean the drain filter thoroughly under running water.
4. Screw the drain filter onto the filter housing clockwise using the filter
key.
NOTICE
Tighten the drain filter well without using force.
5. Screw the lid of the filter housing back on clockwise with the filter key.
NOTICE
Tighten the filter housing well without using force.
6.8
Removing and cleaning the drain tank
Frequency:
If using manual water: Weekly or every 50 cycles (whichever comes first)
or when it is full.
If using automatic water: Daily or every 50 cycles (whichever comes first)
or when it is full.
NOTICE
Do not start a cycle if the DAC Universal is not connected to the drain
tank or drainage system.
✔
The DAC Universal is switched off and cooled down.
✔
The drain tank is cooled down.
✔
You are wearing your protective gear.
Sirona Dental A/S
Operating Instructions DAC Universal
66 15 764 D3472
D3472.201.21.01.02 05.2017