Information for Replacing 4278A with 4288A
Example of Replacing Major Functions (comparison of programs)

Reading out the measurement result

Continuous data output (spew-out mode)
The 4288A does not provide a continuous data output function (spew-out mode). However,
as shown in the following figure, by executing the :READ? command and then supplying
a trigger when executing measurement, you can read out a measurement result after
completion of measurement that is similar to the as-is data output of the 4278A.
Table B-20
Comparison of programs used to read out measurement results
10
REAL Pri (1:5),Sec (1:5)
20
ASSIGN @Agt4278a TO 717
30
OUTPUT @Agt4278a;"DBIN0"
40
!
50
!
60
OUTPUT @Agt4278a;"TRIG2"
70
OUTPUT @Agt4278a;"DST"
80
FOR I=1 TO 5
90
!
100
PRINT "Wait for External Trigger!!"
110
ENTER @Agt4278a;Pri (I),Sec (I)
120
NEXT I
130
OUTPUT @Agt4278a;"DEND"
140
!
150
PRINT "
Pri
","
160
FOR I=1 TO 5
170
PRINT Pri (I),Sec (I)
180
NEXT I
190
!
200
END
234
4278A
Sec
"
4288A
10
REAL Pri (1:5),Sec (1:5)
20
ASSIGN @Agt4288a TO 717
30
OUTPUT @Agt4288a;":CALC:COMP OFF"
40
!
50
OUTPUT @Agt4288a;":ABOR"
60
OUTPUT @Agt4288a;":TRIG:SOUR EXT"
70
!
80
FOR I=1 TO 5
90
OUTPUT @Agt4288a;":READ?"
100
PRINT "Wait for External Trigger!!"
110
ENTER @Agt4288a;Dummy,Pri (I),Sec (I)
120
NEXT I
130
!
140
!
150
PRINT "
Pri
","
160
FOR I=1 TO 5
170
PRINT Pri (I),Sec (I)
180
NEXT I
190
!
200
END
Sec
"
Appendix B