:CORR:COLL

Syntax
[:SENSe]:CORRection:COLLect[:ACQuire] {STANdard1|STANdard2|STANdard3}
Description
Measures the compensation data for OPEN/SHORT/LOAD compensation and turns on the
compensation function.
These compensation data are measured for the measurement frequency used in executing
the command (set with the :SOUR:FREQ command).
When using the multi-compensation function (set to ON with the :CORR:MULT
command), this is measured as the compensation data for the selected channel when
executing the command (selected with the :CORR:MULT:CHAN command). (No query)
Parameters
STANdard1
STANdard2
STANdard3
Related commands :CORR:OPEN on page 165
:CORR:SHOR on page 165
:CORR:LOAD on page 160
:CORR:MULT on page 161
:CORR:MULT:CHAN on page 161
Equivalent key
[Open]
sequence
[Short]
[Load]
Chapter 10
Specifies the OPEN compensation.
Specifies the SHORT compensation.
Specifies the LOAD compensation.
(
[Blue]
-
[4]
) -
OpenMeas
(
[Blue]
-
[5]
) -
ShortMeas
(
-
) -
[Blue]
[6]
LoadMeas
Command Reference
:CORR:COLL
Description
157