Agilent Technologies 4288A Programming Manual page 119

1khz/1mhz capacitance meter
Table of Contents

Advertisement

620
NEXT Ch
630
PRINT "##### Short Data Measurement #####"
640
FOR Ch=0 TO Max_chan
650
OUTPUT @Agt4288a;":CORR:MULT:CHAN ";Ch
660
PRINT "Set the Scanner's Channel to No."&VAL$(Ch)&", and"
670
Result=FNCompen(@Agt4288a,"Short",20)
680
IF Result<>0 THEN Prog_end
690
NEXT Ch
700
PRINT "##### Load Data Measurement #####"
710
FOR Ch=0 TO Max_chan
720
OUTPUT @Agt4288a;":CORR:MULT:CHAN ";Ch
730
PRINT "Set the Scanner's Channel to No."&VAL$(Ch)&", and"
740
Result=FNCompen(@Agt4288a,"Load",.2)
750
IF Result<>0 THEN Prog_end
760
NEXT Ch
770
!
780
OUTPUT @Agt4288a;":RANG ";Nom
790
!
800
! Comparator Setting
810
!
820
OUTPUT @Agt4288a;":CALC:COMP:MODE "&Mode$
830
OUTPUT @Agt4288a;":CALC:COMP:PRIM:NOM ";Nom
840
FOR I=1 TO 3
850
OUTPUT @Agt4288a;":CALC:COMP:PRIM:BIN"&VAL$(I)&" ";L_lim(I);","
;U_lim(I)
860
OUTPUT @Agt4288a;":CALC:COMP:PRIM:BIN"&VAL$(I)&":STAT ON"
870
NEXT I
880
FOR I=4 TO 9
890
OUTPUT @Agt4288a;":CALC:COMP:PRIM:BIN"&VAL$(I)&":STAT OFF"
900
NEXT I
910
OUTPUT @Agt4288a;":CALC:COMP:SEC:LIM ";L_lim_sec;",";U_lim_sec
920
OUTPUT @Agt4288a;":CALC:COMP:SEC:STAT ON"
930
!
940
OUTPUT @Agt4288a;":CALC:COMP:AUXB "&Aux$
950
OUTPUT @Agt4288a;":CREJ "&Low_c_rej$
960
OUTPUT @Agt4288a;":CREJ:LIM ";Low_c_rej_lim
970
OUTPUT @Agt4288a;":CALC:COMP ON"
980
!
990
! Trigger setting
1000
!
1010
OUTPUT @Agt4288a;":TRIG:SOUR EXT"
1020
OUTPUT @Agt4288a;":INIT:CONT ON"
1030
!
1040
! Bin count setting
1050
!
1060
OUTPUT @Agt4288a;":CALC:COMP:COUN ON"
1070
OUTPUT @Agt4288a;":CALC:COMP:COUN:CLE"
1080
OUTPUT @Agt4288a;"*OPC?"
1090
ENTER @Agt4288a;Buff$
1100
!
1110
! SRQ setting
1120
!
1130
OUTPUT @Agt4288a;":STAT:OPER:ENAB 16"
1140
OUTPUT @Agt4288a;"*SRE 128"
1150
!
1160
! Data buffer setting
1170
!
1180
OUTPUT @Agt4288a;":DATA:POIN BUF3,";Max_chan+1
1190
OUTPUT @Agt4288a;":DATA:FEED:CONT BUF3,ALW"
1200
OUTPUT @Agt4288a;":FORM REAL"
1210
!
1220
FOR I=1 TO Max_meas
1230
PRINT "Set DUT of All Channels."
1240
!
Chapter 9
Measurement Applications (Sample Programs)
Measurement with changing channels (scanning)
117