Aver - Agilent Technologies 4288A Programming Manual

1khz/1mhz capacitance meter
Table of Contents

Advertisement

:AVER

Syntax
[:SENSe]:AVERage[:STATe] {ON|OFF|1|0}
[:SENSe]:AVERage[:STATe]?
Description
Turns ON/OFF the averaging function.
Parameters
ON or 1 (initial value)
OFF or 0
Query response
{1|0}<newline><^END>
Related commands :AVER:COUN on page 132
Equivalent key
[Average]
sequence
NOTE
You cannot turn ON/OFF the averaging function directly by using the front panel keys.
Setting the averaging count with the front panel keys will automatically turn on the
function. Note that after the function has started this way you cannot turn it off by using the
front panel keys.
Chapter 10
Turns ON the averaging function.
Turns OFF the averaging function.
(
[Blue]
-
[Meas Time]
)
Command Reference
Description
:AVER
131

Hide quick links: