Syst:test; Syst:vers - Agilent Technologies 4288A Programming Manual

1khz/1mhz capacitance meter
Table of Contents

Advertisement

Command Reference

:SYST:TEST?

:SYST:TEST?
Syntax
:SYSTem:TEST?
Description
Executes the external test and reads out its result. (Query only)
NOTE
Before executing the command, connect the 42090A to the UNKNOWN terminal of the
4288A (or directly connect between L
BNC cables).
For more information on the external test, refer to 4288A Service Manual.
Query response
{numeric 1},{numeric 2},{numeric 3},{numeric 4},{numeric 5}<newline><^END>
For a test item that fails, the following information is read out. When all of the test items
pass, all read-out values are 0.
{numeric 1}:
{numeric 2}:
{numeric 3}:
{numeric 4}:
{numeric 5}:
Related commands *TST? on page 129
Equivalent key
[Config]
sequence

:SYST:VERS?

Syntax
:SYSTem:VERSion?
Description
Returns the SCPI version number of the instrument. (Query only)
Query response
{string}<newline><^END>
The readout data format is YYYY.V, where YYYY indicates the version year and V
indicates the version number within that year.
Equivalent key
No equivalent key is available on the front panel.
sequence
194
Test number (an integer ranging from 1 to 4 as shown below)
1: Entire analog circuitry
2: Signal part (signal level)
3: Signal part (frequency)
4: Measurement part
Setting number for each test (an integer ranging from 1 to 10. Refer to
4288A Service Manual.)
Measurement item (an integer between 1 and 3 as shown below)
1: Absolute value of impedance
2: Voltage level of measurement signal
3: Current level of measurement signal
Expected value of measurement item
Measured value of measurement item
(
[Blue]
-
[-]
) -
Test
-
External
and L
and between H
CUR
POT
-
Yes
and H
with
CUR
POT
Chapter 10