:CALC:COMP:COUN:DATA?

:CALC:COMP:COUN:DATA?
Syntax
:CALCulate:COMParator:COUNt:DATA?
Description
Reads out each count value of BIN1 to BIN9, OUT_OF_BINS, and AUX_BIN of the BIN
counter function.
Reads out all BIN count values regardless of the ON/OFF state of each BIN (set with the
:CALC:COMP:PRIM:BIN{1-9}:STAT command). (Query only)
Query response
{numeric 1},...,{numeric 11}<newline><^END>
Related commands :CALC:COMP:COUN:OVLD? on page 140
:CALC:COMP:COUN on page 137
:CALC:COMP:COUN:CLE on page 137
:CALC:COMP:COUN:MULT:DATA? on page 139
:CALC:COMP on page 134
:CALC:COMP:PRIM:BIN{1-9}:STAT on page 143
Equivalent key
No equivalent key is available on the front panel.
sequence
138
{numeric 1}
The count value of BIN1.
{numeric 2}
The count value of BIN2.
{numeric 3}
The count value of BIN3.
{numeric 4}
The count value of BIN4.
{numeric 5}
The count value of BIN5.
{numeric 6}
The count value of BIN6.
{numeric 7}
The count value of BIN7.
{numeric 8}
The count value of BIN8.
{numeric 9}
The count value of BIN9.
{numeric 10}
The count value of OUT_OF_BINS.
{numeric 11}
The count value of AUX_BIN.