Command Reference
:DATA
:DATA {REF1|REF2}
Syntax
:DATA[:DATA] {REF1|REF2},<numeric>
:DATA[:DATA]? {REF1|REF2}
Description
Sets the reference value used in the deviation measurement mode for the primary
parameter or secondary parameter. This reference value is used when the deviation
measurement mode has been enabled for the primary parameter or secondary parameter
(ON has been specified with the :CALC1:MATH:STAT command or
:CALC2:MATH:STAT command).
The transfer format of data read out with this command conforms to the setup made with
the :FORM command.
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
MAX or MIN can be used to specify the parameters.
Query response
{numeric}<newline><^END>
Related commands :CALC1:MATH:STAT on page 149
:CALC2:MATH:STAT on page 152
:FORM on page 182
Equivalent key
No equivalent key is available on the front panel.
sequence
172
REF1
Sets or reads out the reference value for the primary parameter
used in the deviation measurement mode.
REF2
Sets or reads out the reference value for the secondary parameter
used in the deviation measurement mode.
Description
Reference value used in the deviation measurement mode
Range
-999.99 to 999.99 (for REF1)
-99.999E9 to 99.999E9 (for REF2)
Initial value
0
Unit
Depends on the type of parameter.
Description
<numeric>
Chapter 10