:CALC1:FORM

Syntax
:CALCulate1:FORMat {CP|CS}
:CALCulate1:FORMat?
Description
Specifies the primary parameter to be measured.
When the secondary parameter has been set to G or RP and the primary parameter is set to
CS, the secondary parameter is automatically changed to D. Also, when the secondary
parameter has been set to RS and the primary parameter is set to CP, the secondary
parameter is automatically changed to D.
Parameters
CP (initial value)
CS
Query response
{CP|CS}<newline><^END>
Related commands :CALC2:FORM on page 150
Equivalent key
[Meas Prmtr]
sequence

:CALC1:MATH:EXPR:CAT?

Syntax
:CALCulate1:MATH:EXPRession:CATalog?
Description
Reads out available parameters used when executing the command to specify the
expression of the primary parameter in the deviation measurement mode (the
(:CALC1:MATH:EXPR:NAME command). The query response is always DEV,PCNT.
(Query only)
Query response
DEV,PCNT<newline><^END> (fixed)
Related commands :CALC1:MATH:EXPR:NAME on page 148
Equivalent key
No equivalent key is available on the front panel.
sequence
Chapter 10
Description
Specifies the capacitance value measured using the parallel
equivalent circuit model for the primary parameter.
Specifies the capacitance value measured using the series
equivalent circuit model for the primary parameter.
Command Reference
:CALC1:FORM
147