Calc:comp:auxb; Calc:comp:beep - Agilent Technologies 4288A Programming Manual

1khz/1mhz capacitance meter
Table of Contents

Advertisement

:CALC:COMP:AUXB

Syntax
:CALCulate:COMParator:AUXBin {ON|OFF|1|0}
:CALCulate:COMParator:AUXBin?
Description
Turns ON/OFF the AUX_BIN function for sorting of the comparator function.
Depending on the ON/OFF state of the AUX_BIN function, the following difference
occurs in the sorting result when the measurement result of the secondary parameter
exceeds the limit range.
ON:
OFF:
Parameters
ON or 1
OFF or 0 (initial value)
Query response
{1|0}<newline><^END>
Equivalent key
[Comprtr]
sequence

:CALC:COMP:BEEP

Syntax
:CALCulate:COMParator:BEEPer[:STATe] {ON|OFF|1|0}
:CALCulate:COMParator:BEEPer[:STATe]?
Description
Turns ON/OFF the beep output.
If you turn off the beep output, the beep sound is not produced, regardless of the sorting
result of the comparator.
This command has the same function as the :SYST:BEEP:STAT command.
Parameters
ON or 1 (initial value)
OFF or 0
Query response
{1|0}<newline><^END>
Related commands :SYST:BEEP:STAT on page 191
Equivalent key
[Config]
sequence
Chapter 10
Sorted into AUX_BIN if the measurement result of the primary
parameter is within the limit range. Otherwise, sorted into
OUT_OF_BINS.
Always sorted into OUT_OF_BINS.
Turns ON the AUX_BIN function.
Turns OFF the AUX_BIN function.
(
[Blue]
-
[1]
) -
Aux
Turns ON the beep output.
Turns OFF the beep output.
(
-
) -
[Blue]
[-]
Beep
Command Reference
:CALC:COMP:AUXB
Description
Description
135

Hide quick links:

Advertisement

Table of Contents
loading