Calc:comp:mode - Agilent Technologies 4288A Programming Manual

1khz/1mhz capacitance meter
Table of Contents

Advertisement

:CALC:COMP:MODE

Syntax
:CALCulate:COMParator:MODE {ABS|DEV|PCNT}
:CALCulate:COMParator:MODE?
Description
Determines how to specify the limit range of the primary parameter for the comparator
function.
Parameters
ABS
(initial value)
DEV
PCNT
Query response
{ABS|DEV|PCNT}<newline><^END>
Related commands :CALC:COMP:PRIM:NOM on page 144
:CALC:COMP:PRIM:BIN{1-9} on page 142
Equivalent key
[Comprtr]
sequence
Chapter 10
Specifies the limit border value in an absolute value (absolute mode).
Specifies the border value relative to the reference value
absolute value (absolute tolerance mode).
Specifies the border value relative to the reference value as a
percentage of the reference value
*1.Use the :CALC:COMP:PRIM:NOM command to set the reference value.
(
[Blue]
-
[1]
) -
Limit
-
Description
*1
(percent tolerance mode).
∆ Mode
Command Reference
:CALC:COMP:MODE
*1
as an
141