Preparation for Accurate Measurement (Executing Compensation)

Executing Offset Compensation

Turning ON/OFF the compensation function

When you turn ON the offset compensation, assuming that the measured value before
compensation is
and the offset compensation data is
Meas
compensated as
Meas Offset
To turn ON/OFF the offset compensation, use the following command.
:CORR:OFFS on page 163
You cannot turn ON/OFF the primary parameter and secondary parameter separately.
However, if you set the compensation value to 0, the state is actually the same as OFF even
if the offset compensation is ON. Therefore, in practice you can realize separate ON/OFF
states by setting the compensation value for either parameter to 0.

Setting up the compensation data

To set up the offset compensation data, use the following command.
:CORR:OFFS:DATA on page 164
The entered value is set as the offset compensation data for the measurement frequency at
the time of the entry.
Chapter 4
Executing Offset Compensation
.
, the measured value is
Offset
45