Turning On/Off Aux Bin Function; Setting Up Condition To Make A Beep Sound - Agilent Technologies 4288A Programming Manual

1khz/1mhz capacitance meter
Table of Contents

Advertisement

Sorting Based on Measured Result (Comparator Function)
Setting Up Comparator Function

Turning ON/OFF AUX BIN function

To turn ON/OFF the AUX BIN function, use the following command.
Depending on the ON/OFF state of the AUX BIN function, the sorting result when the
value of the secondary parameter exceeds the limit range varies as shown in Table 7-1.
Table 7-1
Sorting result when measured secondary parameter value exceeds limit range

Setting up condition to make a beep sound

You can select the condition that produces a beep sound based on the comparator sorting
judgment result from the following:
To set the condition that makes a beep sound, use the following command.
You can also disable the beep sound. To turn ON/OFF the beep, use either of the following
commands (they have the same function).
84
:CALC:COMP:AUXB on page 135
Primary parameter sorting result
One of BIN1 to BIN9
Not sorted to any BIN
Beep when the sorting judgment result is OUT_OF_BINS or AUX_BIN
Beep when the sorting judgment result is BIN1 to BIN9
:CALC:COMP:BEEP:COND on page 136
:CALC:COMP:BEEP on page 135
:SYST:BEEP:STAT on page 191
AUX BIN function
OFF
ON
Not applicable
Sorting result
OUT_OF_BINS
AUX_BIN
OUT_OF_BINS
Chapter 7

Hide quick links: