Calc1:Math:expr:name - Agilent Technologies 4288A Programming Manual

1khz/1mhz capacitance meter
Table of Contents

Advertisement

Command Reference

:CALC1:MATH:EXPR:NAME

:CALC1:MATH:EXPR:NAME
Syntax
:CALCulate1:MATH:EXPRession:NAME {DEV|PCNT}
:CALCulate1:MATH:EXPRession:NAME?
Description
Specifies the expression of the primary parameter used when displaying the measurement
result in the deviation measurement mode.
Parameters
Query response
{DEV|PCNT}<newline><^END>
Related commands :CALC1:MATH:STAT on page 149
:CALC2:MATH:EXPR:NAME on page 151
:DATA {REF1|REF2} on page 172
Equivalent key
[∆ Mode]
sequence
148
DEV
Displays the result in the difference between the measurement
(initial value)
value and the reference value
value).
PCNT
Displays the difference between the measurement value and the
reference value
*1.Use the :DATA {REF1|REF2} command to set the reference value.
*2.(Measurement value - reference value) / reference value× 100
(
[Blue]
-
[Meas Prmtr]
Description
*1
(measurement value - reference
*1
*2
in a percentage
) -
Pri
to the reference value.
Chapter 10

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents