:CORR:OPEN

Syntax
[:SENSe]:CORRection:OPEN[:STATe] {ON|OFF|1|0}
[:SENSe]:CORRection:OPEN[:STATe]?
Description
Turns ON/OFF the OPEN compensation.
With the OPEN compensation set to ON, if you change the cable length (set with the
:CAL:CABL command) or frequency shift (set with the :SYST:FSH command), the
OPEN compensation is automatically changed to OFF.
Parameters
ON or 1
OFF or 0 (initial value)
Query response
{1|0}<newline><^END>
Related commands :CAL:CABL on page 133
:SYST:FSH on page 192
:CORR:COLL on page 157
Equivalent key
[Open]
sequence

:CORR:SHOR

Syntax
[:SENSe]:CORRection:SHORt[:STATe] {ON|OFF|1|0}
[:SENSe]:CORRection:SHORt[:STATe]?
Description
Turns ON/OFF the SHORT compensation.
With the SHORT compensation set to ON, if you change the cable length (set with the
:CAL:CABL command) or frequency shift (set with the :SYST:FSH command), the
SHORT compensation is automatically changed to OFF.
Parameters
ON or 1
OFF or 0 (initial value)
Query response
{1|0}<newline><^END>
Related commands :CAL:CABL on page 133
:SYST:FSH on page 192
:CORR:COLL on page 157
Equivalent key
[Short]
sequence
Chapter 10
Turns ON the OPEN compensation.
Turns OFF the OPEN compensation.
(
-
) -
[Blue]
[4]
On/Off
Turns ON the SHORT compensation.
Turns OFF the SHORT compensation.
(
[Blue]
-
[5]
) -
On/Off
Command Reference