Data:poin - Agilent Technologies 4288A Programming Manual

1khz/1mhz capacitance meter
Table of Contents

Advertisement

:DATA:POIN

Syntax
:DATA:POINts {BUF1|BUF2|BUF3},<numeric>
:DATA:POINts? {BUF1|BUF2|BUF3}
Description
Specifies the size of data buffer 1, data buffer 2, or data buffer 3 in number of
measurements. Executing this command returns the pointer to the specified data buffer (the
location to feed measurement data) back to the start.
Parameters
BUF1
BUF2
BUF3
Description
Range
Initial value
Resolution
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
MAX or MIN can be used to specify the parameters.
Query response
{numeric}<newline><^END>
Related commands :DATA:FEED on page 173
:DATA:FEED:CONT on page 174
Equivalent key
No equivalent key is available on the front panel.
sequence
Chapter 10
Description
Specifies data buffer 1 as the buffer to which the <numeric> setup
is applied or reads out the setup of data buffer 1.
Specifies data buffer 2 as the buffer to which the <numeric> setup
is applied or reads out the setup of data buffer 2.
Specifies data buffer 3 as the buffer to which the <numeric> setup
is applied or reads out the setup of data buffer 3.
<numeric>
Number of measurements for the data buffer specified with
{BUF1|BUF2|BUF3}
Data buffer 1: 1 to 200
Data buffer 2: 1 to 200
Data buffer 3: 1 to 1000
Data buffer 1: 200
Data buffer 2: 200
Data buffer 3: 1000
1
Command Reference
:DATA:POIN
175