Reading Out Measured Result - Agilent Technologies 4288A Programming Manual

1khz/1mhz capacitance meter
Table of Contents

Advertisement

Reading Out Measured Result

This section describes how to read out the measured result.
You can read out the measured result in two ways: reading out data for each measurement
or reading out data of several measurements in batch.
The three commands shown in the table below can be used to read out the measured data
for each measurement.
Using the *TRG
command
Using the :FETC?
command
Using the :READ?
command
When you want to read out the data of several measurements in batch, use the data buffer.
Chapter 6
Reading Out Measured Result
Reading Out Measured Result
Available trigger mode
GPIB trigger
(Bus)
All
External trigger (Ext)
Internal trigger (Int)
Readout procedure
Executing *TRG
Readout
Triggering the instrument
Executing :FETC?
Readout
Executing :READ?
Triggering the instrument
Readout
61