Sour:volt - Agilent Technologies 4288A Programming Manual

1khz/1mhz capacitance meter
Table of Contents

Advertisement

Command Reference

:SOUR:VOLT

:SOUR:VOLT
Syntax
:SOURce:VOLTage[:LEVel][:IMMediate][:AMPLitude] <numeric>[MV|M|V]
:SOURce:VOLTage[:LEVel][:IMMediate][:AMPLitude]?
Description
Sets the measurement signal level.
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set. Note that a fraction below the resolution is rounded off.
MAX or MIN can be used to specify the parameters.
Query response
{numeric}<newline><^END>
Equivalent key
[Level]
sequence
188
Description
Measurement signal level
Range
100E-3 to 1
Initial value
1
Unit
V
Resolution
0.1
<numeric>