Agilent Technologies 4288A Programming Manual page 43

1khz/1mhz capacitance meter
Table of Contents

Advertisement

Preparation for Accurate Measurement (Executing Compensation)
390
CASE "Short"
400
Std$="STAN2"
410
CASE "Load"
420
Std$="STAN3"
430
CALL Inp_data("Load(Cp) Value @1kHz",Load1(1))
440
CALL Inp_data("Load(D)
450
CALL Inp_data("Load(Cp) Value @1MHz",Load1(2))
460
CALL Inp_data("Load(D)
470
OUTPUT @Agt4288a;":CORR:CKIT:STAN3:FORM "&Load_para$
480
FOR I=1 TO 2
490
OUTPUT @Agt4288a;":SOUR:FREQ ";Freq(I)
500
OUTPUT @Agt4288a;":CORR:CKIT:STAN3 ";Load1(I);",";Load2(I)
510
NEXT I
520
END SELECT
530 Compen_meas:
!
540
PRINT "Set "&Standard$&"-Connection."
550
INPUT "OK? [Y/N]",Inp_char$
560
IF UPC$(Inp_char$)="Y" THEN
570
FOR I=1 TO 2
580
PRINT "Frequency:";Freq(I)
590
OUTPUT @Agt4288a;":SOUR:FREQ ";Freq(I)
600
OUTPUT @Agt4288a;"*CLS"
610
OUTPUT @Agt4288a;"*OPC?"
620
ENTER @Agt4288a;Buff$
630
! Measurement
640
ON INTR 7 GOTO Meas_end
650
ENABLE INTR 7;2
660
OUTPUT @Agt4288a;":CORR:COLL "&Std$
670 Meas_wait:
GOTO Meas_wait
680 Meas_end:
OFF INTR 7
690
! Error Check
700
OUTPUT @Agt4288a;":SYST:ERR?"
710
ENTER @Agt4288a;Err_no,Err$
720
IF Err_no<>0 THEN
730
PRINT "Error: "&Err$
740
GOTO Compen_meas
750
END IF
760
! Data Check
770
OUTPUT @Agt4288a;":CORR:DATA? "&Std$
780
ENTER @Agt4288a;Para1,Para2
790
Err_flag=0
800
SELECT Standard$
810
CASE "Open"
820
Ym=SQRT(Para1*Para1+Para2*Para2)
830
PRINT "G =";Para1,"B =";Para2,"|Y| =";Ym
840
IF Ym>=Limit THEN Err_flag=1
850
CASE "Short"
860
Zm=SQRT(Para1*Para1+Para2*Para2)
870
PRINT "R =";Para1,"X =";Para2,"|Z| =";Zm
880
IF Zm>=Limit THEN Err_flag=1
890
CASE "Load"
900
OUTPUT @Agt4288a;":CORR:CKIT:STAN3?"
910
ENTER @Agt4288a;Cpref,Dref
920
Bref=2*PI*Freq(I)*Cpref
930
Gref=Bref*Dref
940
Zref=1/SQRT(Gref*Gref+Bref*Bref)
950
Bm=2*PI*Freq(I)*Para1
960
Gm=Bm*Para2
970
Zm=1/SQRT(Gm*Gm+Bm*Bm)
980
PRINT "Cpref=";Cpref,"Dref=";Dref,"|Zref| =";Zref
990
PRINT "Cp
1000
IF ABS((Zm-Zref)/Zref)>=Limit THEN Err_flag=1
1010
END SELECT
1020
IF Err_flag<>0 THEN
Chapter 4
Executing OPEN/SHORT/LOAD Compensation
Value @1kHz",Load2(1))
Value @1MHz",Load2(2))
=";Para1,"D
=";Para2,"|Z|
=";Zm
41