:CORR:CKIT:STAN3:FORM

:CORR:CKIT:STAN3:FORM
Syntax
[:SENSe]:CORRection:CKIT:STANdard3:FORMat
[:SENSe]:CORRection:CKIT:STANdard3:FORMat?
Description
Sets the types of parameters used to define the standard for LOAD compensation.
Parameters
For details on CP and CS, refer to :CALC1:FORM on page 147. For details on D, Q, RP,
and RS, refer to :CALC2:FORM on page 150.
Query response
{CPD|CPQ|CPG|CPRP|CSD|CSQ|CSRS}<newline><^END>
Related commands :CORR:CKIT:STAN3 on page 155
Equivalent key
[Load]
sequence
156
CPD
Specifies CP as the primary parameter, D as the secondary parameter.
(initial value)
CPQ
Specifies CP as the primary parameter, Q as the secondary parameter.
CPG
Specifies CP as the primary parameter, G as the secondary parameter.
CPRP
Specifies CP as the primary parameter, RP as the secondary parameter.
CSD
Specifies CS as the primary parameter, D as the secondary parameter.
CSQ
Specifies CS as the primary parameter, Q as the secondary parameter.
CSRS
Specifies CS as the primary parameter, RS as the secondary parameter.
(