Disp:text1:Fmsd - Agilent Technologies 4288A Programming Manual

1khz/1mhz capacitance meter
Table of Contents

Advertisement

:DISP:TEXT1:FMSD

Syntax
:DISPlay[:WINDow]:TEXT1[:DATA]:FMSD[:STATe] {ON|OFF|1|0}
:DISPlay[:WINDow]:TEXT1[:DATA]:FMSD[:STATe]?
Description
Determines whether to use the fixed point display when displaying both the primary and
secondary parameters. In the following cases, the fixed point display is always used
regardless of this setting (the value of the highest digit is also fixed).
When displaying deviation as a percentage in the deviation measurement mode.
When displaying the results of D and Q of the secondary parameter.
Parameters
ON or 1
OFF or 0 (initial value)
Query response
{1|0}<newline><^END>
Related commands :DISP:TEXT1:FMSD:DATA on page 178
:CALC1:MATH:EXPR:NAME on page 148
:CALC1:MATH:STAT on page 149
:CALC2:MATH:EXPR:NAME on page 151
:CALC2:MATH:STAT on page 152
Equivalent key
[Disp Mode]
sequence
Chapter 10
Specifies the fixed point display.
Specifies the floating point display.
(
[Blue]
-
[Freq]
) -
FixMsd
-
Command Reference
:DISP:TEXT1:FMSD
Description
On/Off
177

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents