:CREJ

:CREJ
Syntax
[:SENSe][:FIMPedance]:CREJect[:STATe] {ON|OFF|1|0}
[:SENSe][:FIMPedance]:CREJect[:STATe]?
Description
Enables/disables the Low C reject function.
When you enable the Low C reject function, if the measured value of the primary
parameter (Cp or Cs) is too small (equal to or less than the boundary value specified with
the :CREJ:LIM command), Low C is detected. For information on the screen display,
GPIB output, and so on when Low C is detected, refer to Appendix E, "At-a-glance Table
of Operations When Overload or Low C is Detected,".
Parameters
Query response
{1|0}<newline><^END>
Related commands :CREJ:LIM on page 167
Equivalent key
[Low C Rej]
sequence
166
ON or 1
Enables the Low C reject function.
OFF or 0
Disables the Low C reject function.
(initial value)
key (Blue key,
Description
key)
[Level]