Agilent Technologies 4288A Programming Manual page 173

1khz/1mhz capacitance meter
Table of Contents

Advertisement

:DATA? {IMON|VMON}
Syntax
:DATA[:DATA]? {IMON|VMON}
Description
Reads out the current monitor value or voltage monitor value of the measured signal. If the
monitor function is disabled (OFF has been specified with the :CALC3:MATH:STAT
command or :CALC4:MATH:STAT command) or the measurement fails, 9.9E37 is read
out. (Query only)
The transfer format of data read out with this command conforms to the setup made with
the :FORM command.
Parameters
IMON
VMON
Query response
{numeric}<newline><^END>
Related commands :CALC3:MATH:STAT on page 153
:CALC4:MATH:STAT on page 153
:FORM on page 182
Equivalent key
No equivalent key is available on the front panel.
sequence
Chapter 10
Description
Reads out the current monitor.
Reads out the voltage monitor.
Command Reference