Agilent Technologies 4288A Programming Manual page 121

1khz/1mhz capacitance meter
Table of Contents

Advertisement

1880
Std$="STAN2"
1890
CASE "Load"
1900
Std$="STAN3"
1910
CALL Inp_data("Load(Cp) Value @1kHz",Load1(1))
1920
CALL Inp_data("Load(D)
1930
CALL Inp_data("Load(Cp) Value @1MHz",Load1(2))
1940
CALL Inp_data("Load(D)
1950
OUTPUT @Agt4288a;":CORR:CKIT:STAN3:FORM "&Load_para$
1960
FOR I=1 TO 2
1970
OUTPUT @Agt4288a;":SOUR:FREQ ";Freq(I)
1980
OUTPUT @Agt4288a;":CORR:CKIT:STAN3 ";Load1(I);",";Load2(I)
1990
NEXT I
2000
END SELECT
2010 Compen_meas: !
2020
PRINT "Set "&Standard$&"-Connection."
2030
INPUT "OK? [Y/N]",Inp_char$
2040
IF UPC$(Inp_char$)="Y" THEN
2050
FOR I=1 TO 2
2060
PRINT "Frequency:";Freq(I)
2070
OUTPUT @Agt4288a;":SOUR:FREQ ";Freq(I)
2080
OUTPUT @Agt4288a;"*CLS"
2090
OUTPUT @Agt4288a;"*OPC?"
2100
ENTER @Agt4288a;Buff$
2110
! Measurement
2120
ON INTR 7 GOTO Meas_end
2130
ENABLE INTR 7;2
2140
OUTPUT @Agt4288a;":CORR:COLL "&Std$
2150 Meas_wait: GOTO Meas_wait
2160 Meas_end: OFF INTR 7
2170
! Error Check
2180
OUTPUT @Agt4288a;":SYST:ERR?"
2190
ENTER @Agt4288a;Err_no,Err$
2200
IF Err_no<>0 THEN
2210
PRINT "Error: "&Err$
2220
GOTO Compen_meas
2230
END IF
2240
! Data Check
2250
OUTPUT @Agt4288a;":CORR:DATA? "&Std$
2260
ENTER @Agt4288a;Para1,Para2
2270
Err_flag=0
2280
SELECT Standard$
2290
CASE "Open"
2300
Ym=SQRT(Para1*Para1+Para2*Para2)
2310
PRINT "G =";Para1,"B =";Para2,"|Y| =";Ym
2320
IF Ym>=Limit THEN Err_flag=1
2330
CASE "Short"
2340
Zm=SQRT(Para1*Para1+Para2*Para2)
2350
PRINT "R =";Para1,"X =";Para2,"|Z| =";Zm
2360
IF Zm>=Limit THEN Err_flag=1
2370
CASE "Load"
2380
OUTPUT @Agt4288a;":CORR:CKIT:STAN3?"
2390
ENTER @Agt4288a;Cpref,Dref
2400
Bref=2*PI*Freq(I)*Cpref
2410
Gref=Bref*Dref
2420
Zref=1/SQRT(Gref*Gref+Bref*Bref)
2430
Bm=2*PI*Freq(I)*Para1
2440
Gm=Bm*Para2
2450
Zm=1/SQRT(Gm*Gm+Bm*Bm)
2460
PRINT "Cpref=";Cpref,"Dref=";Dref,"|Zref| =";Zref
2470
PRINT "Cp
2480
IF ABS((Zm-Zref)/Zref)>=Limit THEN Err_flag=1
2490
END SELECT
2500
IF Err_flag<>0 THEN
2510
PRINT "Out of limit!!"
Chapter 9
Measurement Applications (Sample Programs)
Measurement with changing channels (scanning)
Value @1kHz",Load2(1))
Value @1MHz",Load2(2))
=";Para1,"D
=";Para2,"|Z|
=";Zm
119