Agilent Technologies 4288A Programming Manual page 111

1khz/1mhz capacitance meter
Table of Contents

Advertisement

160
Cbl=1
170
!
180
Mode$="PCNT"
190
Nom=1.E-9
200
!-- [BIN1] --
210
L_lim(1)=-1.0
220
U_lim(1)=1.0
230
!-- [BIN2] --
240
L_lim(2)=-2.0
250
U_lim(2)=2.0
260
!-- [BIN3] --
270
L_lim(3)=-3.0
280
U_lim(3)=3.0
290
!-- [Sec] --
300
L_lim_sec=0.
310
U_lim_sec=.1
320
Aux$="ON"
330
Low_c_rej$="ON"
340
Low_c_rej_lim=10.0
350
!
360
OUTPUT @Agt4288a;"*RST"
370
OUTPUT @Agt4288a;":CALC1:FORM ";Pri$
380
OUTPUT @Agt4288a;":CALC2:FORM ";Sec$
390
OUTPUT @Agt4288a;":SOUR:FREQ ";Freq
400
OUTPUT @Agt4288a;":SOUR:VOLT ";Lvl
410
OUTPUT @Agt4288a;":APER "&Meas_time$
420
OUTPUT @Agt4288a;":CAL:CABL ";Cbl
430
!
440
OUTPUT @Agt4288a;":DISP OFF"
450
OUTPUT @Agt4288a;":SYST:BEEP:STAT OFF"
460
OUTPUT @Agt4288a;":SYST:KLOC ON"
470
!
480
! Compensation
490
!
500
Result=FNCompen(@Agt4288a,"Open",.00002)
510
IF Result<>0 THEN Prog_end
520
Result=FNCompen(@Agt4288a,"Short",20)
530
IF Result<>0 THEN Prog_end
540
Result=FNCompen(@Agt4288a,"Load",20)
550
IF Result<>0 THEN Prog_end
560
!
570
OUTPUT @Agt4288a;":RANG ";Nom
580
!
590
! Comparator Setting
600
!
610
OUTPUT @Agt4288a;":CALC:COMP:MODE "&Mode$
620
OUTPUT @Agt4288a;":CALC:COMP:PRIM:NOM ";Nom
630
FOR I=1 TO 3
640
OUTPUT @Agt4288a;":CALC:COMP:PRIM:BIN"&VAL$(I)&" ";L_lim(I);","
;U_lim(I)
650
OUTPUT @Agt4288a;":CALC:COMP:PRIM:BIN"&VAL$(I)&":STAT ON"
660
NEXT I
670
FOR I=4 TO 9
680
OUTPUT @Agt4288a;":CALC:COMP:PRIM:BIN"&VAL$(I)&":STAT OFF"
690
NEXT I
700
OUTPUT @Agt4288a;":CALC:COMP:SEC:LIM ";L_lim_sec;",";U_lim_sec
710
OUTPUT @Agt4288a;":CALC:COMP:SEC:STAT ON"
720
!
730
OUTPUT @Agt4288a;":CALC:COMP:AUXB ON"
740
OUTPUT @Agt4288a;":CREJ "&Low_c_rej$
750
OUTPUT @Agt4288a;":CREJ:LIM ";Low_c_rej_lim
760
OUTPUT @Agt4288a;":CALC:COMP ON"
770
!
780
! Trigger setting
Chapter 9
Measurement Applications (Sample Programs)
Measurement with Auto-sorting System
109

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents