Calc:comp:prim:bin{1-9 - Agilent Technologies 4288A Programming Manual

1khz/1mhz capacitance meter
Table of Contents

Advertisement

Command Reference

:CALC:COMP:PRIM:BIN{1-9}

:CALC:COMP:PRIM:BIN{1-9}
Syntax
:CALCulate:COMParator:PRIMary:BIN{1|2|3|4|5|6|7|8|9} <numeric 1>,<numeric 2>
:CALCulate:COMParator:PRIMary:BIN{1|2|3|4|5|6|7|8|9}?
Description
Sets the limit ranges of BIN1 to BIN9 as the primary parameter used in the comparator
function, depending on the designation method set using the :CALC:COMP:MODE
command.
This command only sets the limit range. To activate it, use the
:CALC:COMP:PRIM:BIN{1-9}:STAT command to turn on the BIN for which the limit
range has been set.
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set. Note that the unit of the parameter may change depending on the
limit range designation method.
MAX or MIN can be used to specify the parameters.
Query response
{numeric 1},{numeric 2}<newline><^END>
Related commands :CALC:COMP:MODE on page 141
:CALC:COMP:PRIM:BIN{1-9}:STAT on page 143
:CALC:COMP:SEC:LIM on page 145
:CALC:COMP on page 134
Equivalent key
[Pri Low]
sequence
[Comprtr]
142
Description
The lower limit value of the
limit range.
Range
-999.999 to 999.999
Initial value
0
Unit
F (farad) or % (percent)
and
[Pri High]
(Only BIN1 can be set.)
(
-
) -
[Blue]
[1]
<numeric 1>
-
(All BINs can be set.)
Limit
Pri
<numeric 2>
The upper limit value of the
limit range.
-999.999 to 999.999
0
F (farad) or % (percent)
Chapter 10